User`s guide
Performance Tests Optional Performance Tests
138 Agilent 81600B Tunable Laser Source Family, Fourth Edition
5
Fix the optical fiber at the input of the FP-Etalon; allow the FP-Etalon to
stabilize for at least 5 minutes.
6 To configure the instruments in all of the subsequent measurements
follow the settings reported below - except if otherwise specified:
a ensure the modulation of the source is turned off
b Set the power and sweep settings:
7 Before taking the measurement:
a perform a lambda zero (via menu) on the TLS module;
b zero the power-meter (make sure the TLS output is disabled).
Reference Scans These scans are executed in stepped mode, one per wavelength range,
and will provide the reference wavelength measurements (once the
threshold-crossing analysis is performed).
Tunable Laser Channel [Menu] Parameters Values
<Wavelength Mode><λ>
<Source State><Off>
<Power Unit><dBm>
<Power Mode><Automatic>
Modulation: <mod src><Off>
Table 12 Power and Sweep settings for dynamic accuracy tests
81600B
#200
81600B
#160
81600B
#150
81600B
#140
81600B
#130
81600B
#142 #003
81600B
#142
81600B
#132
TLS output
port
Low SSE,
Output 1
Low SSE,
Output 1
Low SSE,
Output 1
Low SSE,
Output 1
Low SSE,
Output 1
High
Power,
Output 2
High Power,
Output 2
High Power,
Output 2
TLS output
power
- 2 dBm - 6 dBm - 3 dBm - 7 dBm - 9 dBm +3.5 dBm +5 dBm 0 dBm
PM range 0 dBm 0 dBm 0 dBm 0 dBm 0 dBm +10 dBm +10 dBm 0 dBm
Wavelengt
h step
2.0pm 2.0pm 2.0pm 2.0pm 2.0pm 2.0pm 2.0pm 2.0pm
Wavelengt
h range 1
1475-1485
nm
1510-1514
nm
1520-1524
nm
1420-
1424
nm
1270-1274
nm
1420-1424
nm
1420-1424
nm
1260-1264
nm
Wavelengt
h range 2
1610-1620
nm
1165-1620
nm
1566-1570
nm
1476-
1480
nm
1371-1375
nm
1476-1480
nm
1476-1480
nm
1270-1280
nm