User`s guide
Performance Tests Optional Performance Tests
132 Agilent 81600B Tunable Laser Source Family, Fourth Edition
3
Move to the TLS channel of the 8164A/B mainframe. Set the [Menu]
parameters to:
4 Set the output power of the TLS module to:
5 Set the Optical Spectrum Analyzer:
a Set span to 30 nm.
Press Span, and enter this value.
b Set the Resolution Bandwidth to 1 nm.
Press [AMPL], press [BW Swp] then enter this value.
c Set the Sensitivity to -60 dBm.
Press [AMPL], press [SENS] then enter this value.
6 Record the spectrum (run a single sweep):
a Press Peak Search in the Marker field.
b Set the Marker to Center Wavelength and note its displayed wave-
length as:
OSA_λ_center = ________________ nm
7 Find the maximum power level at OSA_λ_center, peak_power, and
enter the result in the test record in [pW]:
Peak_power = ___________ 10
-12
W = ___________ pW
Tunable Laser Channel [Menu] Parameters Values
<Wavelength Mode><λ>
<Source State><Off>
<Power Unit><dBm>
<Power Mode><Automatic>
Modulation: <mod src><Off>
Table 10 Output Power setting for Total SSE - High Power Output:
Tunable Laser Source module Power (P) Wavelength
Agilent 81600B #200, Output 2 + 8.00 dBm 1570 nm
Agilent 81600B #160, Output 2 + 5.00 dBm 1570 nm
Agilent 81600B #150, Output 2 + 5.00 dBm 1550 nm
Agilent 81600B #140, Output 2 + 5.00 dBm 1460 nm
Agilent 81600B #142 #003 + 6.00 dBm 1460 nm
Agilent 81600B #142 + 7.50 dBm 1460 nm
Agilent 81600B #132 + 7.00 dBm 1330 nm