User`s guide

Optional Performance Tests Performance Tests
Agilent 81600B Tunable Laser Source Family, Fourth Edition 131
Optional Performance Tests
These tests refer to some typical characterics of the TLS that are not
guaranteed, and which are not part of the standard re-calibration.
However, the tests can be performed in qualified Agilent Service Centers
on special request.
Signal-to-Total-Source Spontaneous
Emission Ratio - High Power Output
Note: This test does not apply to 81600B #130
For definition, see “Signal to total source spontaneous emission ratio” on
page 47.
Test Procedure
This optional test procedure is applicable to:
It measures the Total SSE of the module at the outputs indicated.
1 Connect the Tunable Laser module (DUT) to the Optical Spectrum
Analyzer as shown in Figure 40:
Figure 40 Test Setup for Total Source Spontaneous Emission Test - High Power output
2 On the Agilent 81600B #140, #150, #160 and #200, make sure that
Output 2, the High Power output, is connected to the Optical Spectrum
Analyzer.
Agilent 81600B #200, Output 2, High Power
Agilent 81600B #160, Output 2, High Power
Agilent 81600B #150, Output 2, High Power
Agilent 81600B #140, Output 2, High Power
Agilent 81600B #142 #003
Agilent 81600B #142
Agilent 81600B #132