User`s guide

Power Tests Performance Tests
Agilent 81600B Tunable Laser Source Family, Fourth Edition 117
Test Procedure - Low SSE output
Note: This test does not apply to 81600B #130
This test procedure is applicable to:
It measures the SSE of the module at the outputs indicated.
1 Set up the equipment as shown in Figure 35:
Figure 37 Test Setup for Signal-to-Source Spontaneous Emission Test - Low SSE Output.
Agilent 81600B #200, Output 1, Low SSE
Agilent 81600B #160, Output 1, Low SSE
Agilent 81600B #150, Output 1, Low SSE
Agilent 81600B #140, Output 1, Low SSE
Use the correct connector for the Fiber Bragg Grating The Fiber
Bragg Grating must be connected to the TLS module using:
a straight connector, if you are testing a TLS module with option #071
an angled connector, if you are testing a TLS module with option #072
NOTE
Fiber Bragg Grating Wavelength
For Agilent 81600B #200, #160 or #150
use an FBG of: λ
FBG
1520 nm
For Agilent 81600B #140, use an FBG of:
λ
FBG
1407 nm
NOTE