User`s guide
Performance Tests Power Tests
114 Agilent 81600B Tunable Laser Source Family, Fourth Edition
3
Move to the TLS channel of the 8164A/B mainframe. Set the [Menu]
parameters to:
4 Make sure the optical output is switched off.
5 Set the wavelength of the TLS module to
6 Set the power of the TLS to its maximum specified output power (as
given in the Test Record).
7 Switch on the TLS output.
8 Initialize the OSA. Press [Preset] then select <Auto Meas.>.
9 Set the OSA:
•Set Span to 4nm,
Press [Span] then enter the value.
•Set Resolution Bandwidth (RBW) to 0.5 nm,
Press [Ampl], press [BW Sup], then enter the value.
•Set Sensitivity to -60dBm,
Press [Ampl], press [Sens], then enter the value.
•Set Wavelength to the value given at step 5.
Tunable Laser Channel [Menu] Parameters Values
<Wavelength Mode><λ>
<Source State><Off>
<Power Unit><dBm>
<Power Mode><Automatic>
Modulation: <mod src><Off>
TLS module -
High Power output
Wavelength [λ]
Agilent 81600B #200 - Output 2 1440.000 nm
Agilent 81600B #160 - Output 2 1495.000 nm
Agilent 81600B #150 - Output 2 1450.000 nm
Agilent 81600B #140 - Output 2 1370.000 nm
Agilent 81600B #142 #003 1370.000 nm
Agilent 81600B #142 1370.000 nm
Agilent 81600B #132 1260.000 nm