User`s guide

Power Tests Performance Tests
Agilent 81600B Tunable Laser Source Family, Fourth Edition 105
2 Move to the TLS channel of the Agilent 8164A/B Lightwave
Measurement System and set the [Menu] parameters to:
3 For the 81600B #200, #160, #150, #140 and #130, set
<Optical Output> to <High Power(2)>.
4 Set the inital wavelength and power of the TLS to:
5 Connect the fiber to the optical output.
6 Make sure the optical output is switched off.
7 For the 81624A/B power meter channel:
Zero the 81624A/B. Select <Menu> then <Zero>,
•Set the power range manually to -50dBm,
•Set the Averaging Time to 500 ms,
Select <dB> as the power units,
•Set λ, the wavelength, to the same as the TLS module, as given in step 4.
8 Switch on the TLS output.
9 At the 81624A/B, select <Menu> then <Disp Ref>
Tunable Laser Channel [Menu] Parameters Values
<Wavelength Mode><λ>
<Source State><Off>
<Power Unit><dBm>
<Power Mode><Manual Att>
Modulation: <mod src><Off>
TLS module -
High Power output, no attenuation
Wavelength
[λ]
Power [P]
Attenuatio
n [Atten]
Agilent 81600B #200 - Output 2 1440.000 nm - 1.000 dBm 59.000 dB
Agilent 81600B #160 - Output 2 1495.000 nm - 1.000 dBm 59.000 dB
Agilent 81600B #150 - Output 2 1450.000 nm - 1.000 dBm 59.000 dB
Agilent 81600B #140 - Output 2
a
a
For the 81600B #140 and #142 #003, some wavelengths are set to odd values to avoid
conflict with water absorption lines.
1420.200 nm - 3.000 dBm 57.000 dB
Agilent 81600B #142 #003
*
1420.200 nm - 1.500 dBm 58.000 dB
Agilent 81600B #130 - Output 2 1260.200 nm - 3.000 dBm 57.000 dB