User Manual 59
Bottle Number Cables
Examples of instrument specific bottle number cables are included in this
section.
Figure 21 35900–60670, 35900–60920: General use — spade lugs
Connector 1 Connector 2 Remote signal
1BlackGND
2WhitePrepare
3RedStart
4 Green Shut Down
5 Brown Reserved
6BluePower On
7OrangeReady
8YellowStop
9VioletStart Request
- Black tubing Shield
Note: When used for remote control, this cable attaches directly to remote port on back
of 35900. When used for digital I/O or bottle numbers, the digital I/O adaptor cable must
be used to connect to the 35900.