User manual

44 User Manual
6 Parts Replacement
3 Remove the four screws holding the front panel to the unit.
4 Gently pull the front panel forward until it detaches from the frame.
5 Remove the ground wire connection (green wire) from the ground clip by
pulling it firmly.
6 When you are ready to replace the front panel, reattach the green ground
wire to the ground clip on the chassis.
7 Fit the bottom lip of the front panel under the lip of the frame and push it
against the frame, fitting the power switch into the hole.
8 Connect the front panel cable to the connector, pressing firmly until the
ears of the connector lock over the ends of the black connector.
9 Replace the four screws, two on each side.
Remove the MIO Bracket
1 Remove the top cover. (See “Remove/Replace the Top Cover” on page 42.)
2 Remove the MIO card. (See “Connecting to a TCP/IP Network” on page 20.)
3 Remove the three screws attaching the MIO bracket to the frame. (See
Figure 15 on page 51.)
4 Firmly pull up on both the plate and the connector board, holding them
together at a right angle.
Remove the Connector Board
1 Remove the top cover. (See “Remove/Replace the Top Cover” on page 42.)
2 Remove the MIO bracket. (See “Remove the MIO Bracket” on page 44.)
3 Remove the two screws attaching the connector board to the plate. The
board will separate from the plate. (Figure 15 on page 51.)
WARNING
This procedure is intended for use by Agilent-trained service personnel only. There
are no user-serviceable parts inside the 35900.
WARNING
This procedure is intended for use by Agilent-trained service personnel only. There
are no user-serviceable parts inside the 35900.