5 Startup
Pre-Run Checklist
Before using the 35900 with the operating software on the host computer,
make sure you have completed the following:
❏ Rear Panel configuration switches 3 through 6 are DOWN (normal
operation)
❏ MIO card is correctly configured for host computer communications
❏ MIO card connected to host computer or network
❏ Analog signal cables connected
❏ Remote control cables connected (if needed)
❏ Digital I/O cables connected (if needed)
❏ Power cord connected
35900 Power Up
Power up the 35900 by pressing the line switch (lower left corner of the front
panel).
• All front panel indicators should turn on for 1 to 2 seconds.
• All indicators should turn off, except for POWER (on steadily) and FAULT
(blinks while self-tests are performed).
• After approximately 30 seconds, the FAULT indicator should turn off. Some
other indicators may turn on, depending on the instrument configuration.
WARNING
The power cord must be connected to an outlet with a protective earth contact to
ground the unit and minimize shock hazard.