Technical data

30 Understanding Your Agilent ChemStation
1 Agilent ChemStation Features
About the ChemStation Software
All reports have time stamps and traceable page numbering (page x of y
pagination style). The user may select the level of detail in each report ranging
from simple summary reports to complete system details (see the Reporting
section above).
GLP Save register files, specified as part of the method configuration, save all
the original data, including sample information, data analysis method,
chromatographic/electropherographic signals, instrument conditions,
integration and quantification results, report data and the run logbook in one
checksum protected binary file. This is an uneditable binary format that
ensures the originality of the results.The file includes a revisioning scheme
that indicates if data has been reprocessed.
Control sample types may be defined in the sequence table and used to
automatically check the instrument performance against quality control
sample results when the instrument is running unattended. Results that are
outside the user-specified acceptable range will stop the automatic execution
of the instrument.