Technical data

210 Understanding Your Agilent ChemStation
9 Data Review, Reprocessing and Batch Review
Batch Configuration
the compound list contains all compounds found in the method that was
loaded for batch review.
if calibration samples only are displayed in the batch table (samples and
control samples are hidden), the compound table shows additional columns
for calibration-related information (expected amount, relative error and
absolute error).
if control runs only are displayed in the batch table (samples and
calibration samples are hidden), the compound table shows additional
columns for any defined control limits.
For columns containing compound-specific information, you can include the
name of the compound in the table title by adding %s to the column
specification.
Batch Report
The batch report contains two tables that are generally analogous to the batch
table and the compound table; these tables are also user-configurable.
For columns containing compound-specific information, you can include the
name of the compound in the table title by adding %s to the column
specification. Multi-line headers are allowed; you insert the character ‘|’ at the
point where you want the line to break.
User Interface
Batch review provides a choice of two user interfaces:
the standard interface includes a button bar, with buttons mirroring most
of the Batch menu items, together with the batch table and compound table;
a minimal interface provides a button bar similar to the standard interface,
but replaces the batch table and compound table with a combo box that
contains only the information specified for the batch table. The minimal
interface button bar does not contain batch table-related or compound
table-related buttons.