Technical data

Understanding Your Agilent ChemStation 197
Automation
8
Cyclic Recalibration Sequences with Multiple Vials Containing the Same Dilution of a Standard
The execution order is:
Vial 1 (Cal1a)
Vial 5 (Cal2a)
6 injections out of vial 10 (Sample10)
Vial 2 (Cal1b)
Via 6 (Cal2b)
6 injections out of vial 11 (Sample11)
Vial 3 (Cal1c)
Vial 7 (Cal2c)
6 injections out of vial 12 (Sample12)
Vial 1 (Cal1a)
Vial 5 (Cal2a)
6 injections out of vial 13 (Sample13)
Vial 2 (Cal1b)
Vial 6 (Cal2b)
etc.
Table 2 2 Cyclic Recalibration Sequence with 3 Vials Defined for Each Level
Vial No. Sample Name Sample Type Method Name No. of Inj. Lvl Upd RT Upd RF Interval
1 Cal1a Calib MethodA 1 1 Avg Avg 6
2 Cal1b Calib MethodA 1 1 Avg Avg 6
3 Cal1c Calib MethodA 1 1 Avg Avg 6
5 Cal2a Calib MethodA 1 2 Avg Avg 6
6 Cal2b Calib MethodA 1 2 Avg Avg 6
7 Cal2c Calib MethodA 1 2 Avg Avg 6
10 Sample10 Sample MethodA 6
11 Sample11 Sample MethodA 6
12 Sample12 Sample MethodA 6
13 Sample13 Sample MethodA 6
14 Sample14 Sample MethodA 6