Technical data

186 Understanding Your Agilent ChemStation
8 Automation
Cyclic Multiple-Level Calibration Sequences
Cyclic Multiple-Level Calibration Sequences
This type of sequence uses different calibration samples to recalibrate a
multiple-level calibrated method.
The following example describes a two-method sequence comprising method A
and method B to analyze two groups of samples. Both methods are
multiple-level calibration methods that will recalibrate automatically at
defined intervals.
For each method the Sequence Table has three entries:
Two calibration levels:
Sequence lines 1 and 2 in method A.
Sequence lines 8 and 9 in method B.
Five entries for the samples:
Sequence line 3 through 7 in method A.
Sequence line 10 through 14 for method B.
The calibrations are specified at regular intervals by the recalibration interval
entry in the Sequence Recalibration Table.
Method A will recalibrate after every two samples.
Method B will recalibrate after every three samples.
The Sequence Table below is truncated to simplify the example.
Table 1 6 Sequence Table for Method A and Method B
Line Vial Method
Name
Inj/Vial Sample
Type
Cal
Level
Update RF Update RT Interval
1 1 Method A 1 Calibration 1 Average No update 2
2 2 Method A 1 Calibration 2 Average No update 2
310 Method A 1
411 Method A 1
512 Method A 1