Technical data

146 Understanding Your Agilent ChemStation
7 Calibration
Unknown Samples
Unknown Samples
An unknown sample is a sample containing an unknown amount of the
compound to be quantified.
To find out how much of the compound is in the unknown sample, you must:
create a calibration curve for the compound,
inject an aliquot of your unknown sample and run the analysis in exactly
the same way as for the calibration sample,
determine from the signal the response, which is the area or height of the
peak due to the unknown amount of the compound, and
use the calibration curve to calculate the amount of the compound in the
unknown sample.
For example, if the area of peak in the unknown sample is 500, you can
determine that the amount in the unknown is 5 ng/µl, by using the calibration
curve shown in Figure 33 on page 146.
Figure 33 Signal from Unknown Sample and Calibration Curve