Technical data

Understanding Your Agilent ChemStation 129
Peak Identification
6
Peak Matching Rules
Peak Matching Rules
The following rules apply to the peak matching process:
if a sample peak falls within the peak matching window of a component
peak from the calibration table, the peak is given the attributes of that
component,
if more than one sample peak falls within the peak matching window, then,
the peak closest to the expected retention/migration time is identified as
that component,
if a peak is a time reference or internal standard, then the largest peak in
the window is identified as that component,
if peak qualifiers are also used then the peak ratio is used in combination
with the peak matching window to identify the component peak,
if the peak is a qualifier peak, the measured peak closest to the main peak
of the compound is identified, and
if a sample peak does not fall in any peak matching window, it is listed as an
unknown component.