Installation guide
Setting up a LAN Instrument Network 9
Instrument Firmware Requirements
Table 1: Instrument Requirements
Instrument Type Model Supported instrument
firmware revision
Method for setting IP
addresses
7890A G3440A A.01.10.3 Set at GC keypad
7820A G4350A 7820A.01.10.013.1.bin
(E,C,J)
7820A.01.10.013.Ru.bin
Set using software
keypad
6890N GC G1530N/
G1540N
N.06.07 for 7693A
And N.05.06
LAN Assembly
Firmware 04.7B3
Set at GC keypad
6890 Plus, 6890A G1530A/
G1540A
A.03.08 Chip Set Determined by Jet
Direct card used – see
below.
6850 Series GC
SN≤US00003200
G2630A
A.03.07 for 7693A
support and A.03.03
Determined by Jet
Direct card used – see
below.
6850 Series GC
SN≥US10243001
G2630A A.06.02 for 7693A and
A.05.04
LAN Assembly
Firmware 04.7B3
Set at GC keypad
35900E 35900E E.01.02 Determined by Jet
Direct card used – see
below.
Table 2: Jet Direct Card Requirements
Product Hardware
Minimum
Rev
Support Statements
G1369B LAN Interface
card
B.01.01 10/100BaseT
Supported with GC and 35900E
Replaces the G1369A card mid 2010
Set IP address with Telnet or BootP
G1369A
LAN Interface
card
C.03.00
10/100BaseT
Supported with GC and 35900E
Set IP address with Telnet or BootP
J4100A
Jet Direct Card
K.08.32
10/100 Base T
Set IP address with Telnet or BootP
J2552A/B
(obsolete)
Jet Direct Card
A.08.32
10 Base T only
Needs BootP Service (BootP must be running if
the GC is cycled)
LAN address is volatile (Upgrade to G1369B)