Installation guide

Setting up a LAN Instrument Network 18
Figure 8: Telnet information
Information about the card
Product id, firmware revision s, MAC address
Initialization mode is Bootp. The connected PC/Bootserver is
134.40.30.184
Active TCP/IP settings
Stored TCP/IP settings in non-volatile memory (not visible if equal to
active TCP/IP settings)
Connected to PC with controller software (e.g. Agilent ChemStation),
here not connected
5. Change the IP address (in this example 134.40.24.230) and type
“/” to list current settings.
Figure 9: Change the IP address
Enter ip 130.40.24.158 [Enter]. Changes the TCP/IP setting.
Enter / [Enter]. It then displays these settings:
Initialization mode is Bootp
The connected PC/Bootserver is 134.40.30.184
Active TCP/IP settings
Stored TCP/IP settings in non-volatile memory
Last user change IP 134.40.24.150 (not active yet, requires mode
“Using Stored” and re-start)
6. When you have finished typing the configuration parameters,
type:
quit and press [Enter] to store the configuration parameters
or
exit and press [Enter] to exit without storing parameters.
If the Initialization Mode Switch is changed now to Using
Stored” mode, the instrument will take the stored settings
when the module is re-booted. In the example above it would
be 134.40.24.158 on QUIT and 134.40.24.160 on EXIT.