Installation guide

Setting up a LAN Instrument Network 15
Initialization Mode Selection
The following initialization modes are selectable:
NOTE
Compared to the G1369A LAN card, SW 7 and SW 8 must always be in OFF position on the
G1369B LAN card, otherwise the selected modes are not working.
Bootp
When the initialization mode “Bootp” is selected, the card tries to
download the parameters from a Bootp Server. The parameters
obtained become the active parameters immediately. They are not
stored to the non-volatile memory of the card. Therefore, the
parameters are lost with the next power cycle of the card.
Bootp & Store
When “Bootp & Store” is selected, the parameters obtained from a
Bootp Server become the active parameters immediately. In
addition, they are stored to the non-volatile memory of the card.
After a power cycle the parameters are still available.
Using Stored
When initialization mode “Using Stored” is selected, the parameters
are taken from the non-volatile memory of the card. The TCP/IP
connection will be established using these parameters. The
parameters were configured previously by one of the methods
described below.
Using Default
When “Using Default” is selected, the factory default parameters are
taken instead. These parameters enable a TCP/IP connection to the
LAN Interface without further configuration.