Chapter 4 91
IEEE Common Commands
Common Command Descriptions
*RST (Reset)
The *RST command resets the Test Set. When the *RST command is
received the majority of fields in the Test Set are “restored” to a default
value, some fields are “maintained” at their current state and some are
“initialized” to a known state. Refer to “Instrument Initialization” on
page 177 for further information. Other operational characteristics are
also affected by the *RST command as follows:
• All pending operations are aborted.
• The Test Set’s display screen is in the UNLOCKED state.
• Measurement triggering is set to TRIG:MODE:SETT FULL;RETR
REP.
• Any previously received Operation Complete command (*OPC) is
cleared.
• Any previously received Operation Complete query command
(*OPC?) is cleared.
• The power-up self-test diagnostics are not performed.
• The contents of the SAVE/RECALL registers are not affected.
• Calibration data is not affected.
• The HP-IB interface is not reset (any pending Service Request is not
cleared).
• All Enable registers are unaffected: Service Request, Standard
Event, Communicate, Hardware #1, Hardware #2, Operation,
Calibration, and Questionable Data/Signal.
• All Negative Transition Filter registers are unaffected:
Communicate, Hardware #1, Hardware #2, Operational, Calibration,
and Questionable Data/Signal.
• All Positive Transition Filter registers are unaffected: Communicate,
Hardware #1, Hardware #2, Operational, Calibration, and
Questionable Data/Signal.
• The contents of the RAM are unaffected.
• The contents of the Output Queue are unaffected.
• The contents of the Error Queue are unaffected.