48 Chapter 1
O:\Manuals\E6380A_Progguid\Book\Usehpib.fm
Using HP-IB
Remote/Local Modes
Remote/Local Modes
Remote Mode
In Remote mode all front-panel keys are disabled (except for the
LOCAL key, POWER switch, Volume control and Squelch control). The
LOCAL key is only disabled by the Local Lockout bus command. When
in Remote mode and addressed to Listen the Test Set responds to the
Data, Remote, Local, Clear (SDC), and Trigger messages. When the
Test Set is in Remote mode, the R annunciator will be displayed in the
upper right corner of the display screen and triggering is set to the state
it was last set to in Remote mode (if no previous setting, the default is
FULL SETTling and REPetitive RETRiggering). When the Test Set is
being addressed to Listen or Talk the L or T annunciators will be
displayed in the upper-right corner of the display screen.
Local Mode
In Local mode the Test Set’s front-panel controls are fully operational.
The Test Set uses FULL SETTling and REPetitive RETRiggering in
Local mode. When the Test Set is being addressed to Listen or Talk the
L or T annunciators will be displayed in the upper-right corner of the
display screen.
Remote or Local Mode
When addressed to Talk in Remote or Local mode, the Test Set can
issue the Data and Status Byte messages and respond to the Take
Control message. In addition the Test Set can issue the Service Request
Message (SRQ). Regardless of whether it is addressed to talk or listen,
the Test Set will respond to the Clear (DCL), Local Lockout, Clear
Lockout/Set Local, and Abort messages.