User`s manual
Chapter 9 321
Error Messages
Numbered HP-IB Error Descriptions
Example 9-109 Error -608
Initialization of Input Module Relay Count file failed.
Indicates that the Test Set was not able to initialize the Input Module
Relay Switch Count EEPROM file during installation of a new input
module. This error is most probably generated as a result of a hardware
error or failure. Refer to the Test Set’s for diagnostic information.
Example 9-110 Error -1301
Handoff attempted while not in Connect state.
Indicates that an attempt was made to handoff a mobile station to a
new voice channel when the Call Processing Subsystem was not in the
Connect state.
Example 9-111 Error -1302
Release attempted while not in Connect state.
Indicates that an attempt was made to send a Release message to a
mobile station when the Call Processing Subsystem was not in the
Connect state.
Example 9-112 Error -1303
Page attempted while not in Active state.
Indicates that an attempt was made to Page a mobile station when the
Call Processing Subsystem was not in the Active state.
Example 9-113 Error -1304
Origination attempted while not in Active state.
Indicates that a mobile station attempted to originate a call to the
simulated Base Station when the Call Processing Subsystem was not in
the Active state.
Example 9-114 Error -1305
Registration attempted while not in Active state.
Indicates that an attempt was made to send a Registration message to
a mobile station when the Call Processing Subsystem was not in the
Active state.
Example 9-115 Error -1306
Origination in progress.
Indicates that an attempt was made to; register, page, handoff, release,
order a change in power level, put the mobile station in maintenance
mode, or send an alert message to the mobile station while an
origination was in progress.