User`s manual

Chapter 9 313
Error Messages
Numbered HP-IB Error Descriptions
Example 9-52 Error -211
Trigger ignored
Indicates that a GET, *TRG, or triggering signal was received and
recognized by the device but was ignored because of device timing
considerations. For example, the device was not ready to respond.
Example 9-53 Error -212
Arm ignored
Indicates that an arming signal was received and recognized by the
device but was ignored.
Example 9-54 Error -213
Init ignored
Indicates that a request for a measurement initiation was ignored as
another measurement was already in progress.
Example 9-55 Error -214
Trigger deadlock
Indicates that the trigger source for the initiation of a measurement is
set to GET and subsequent measurement query is received. The
measurement cannot be started until a GET is received, but the GET
would cause an INTERRUPTED error.
Example 9-56 Error -215
Arm deadlock
Indicates that the arm source for the initiation of a measurement is set
to GET and subsequent measurement query is received. The
measurement cannot be started until a GET is received, but the GET
would cause an INTERRUPTED error.
Example 9-57 Error -220
Parameter error
Indicates that a program data element related error occurred.
Example 9-58 Error -221
Settings conflict
Indicates that a legal program data element was parsed but could not
be executed due to the current device state (see IEEE 488.2, 6.4.5.3 and
11.5.1.1.5).