208 Chapter 7
O:\Manuals\E6380A_Progguid\Book\Mencard.fm
Memory Cards/Mass Storage
Mass Storage Access
Mass Storage Access
Program and data files stored on the Test Set’s various mass storage
locations can be selectively accessed from the following screens:
• The TESTS (IBASIC Controller) screen.
Any type of file can be accessed from this screen, either through an
IBASIC program or the IBASIC command line.
• The SOFTWARE MENU screen using the Select Procedure
Location: and Select Procedure Filename: fields.
Only procedure files shipped with Agilent Technologies software or
procedure files created using the TESTS (Save/Delete Procedure)
screen of the Tests Subsystem can be accessed using these fields.
When created, procedure file names must include the .PRC
extension (FM_TEST.PRC).
A corresponding code file with the extension .PGM (FM_TEST.PGM)
must reside on the same media for the procedure to work.
• The TESTS (Save/Delete Procedure) screen using the Select
Procedure Location: and Enter Procedure Filename: fields.
This screen is used to create “procedure” files. When created,
procedure file names file names must include the .PRC extension
(FM_TEST.PRC).
Save/Recall register files, stored on the Test Set’s various mass storage
locations, can be accessed using the front-panel SAVE and RECALL
keys.