User`s manual

206 Chapter 7
O:\Manuals\E6380A_Progguid\Book\Mencard.fm
Memory Cards/Mass Storage
Default Mass Storage Locations
Default Mass Storage Locations
Built-in IBASIC Controller
Under the following conditions, the default mass storage location for
the built-in IBASIC controller is the front-panel card slot (mass storage
volume specifier ":INTERNAL,4"):
power-up
initializing RAM with the SERVICE screen’s RAM Initialize
function
formatting an SRAM card using the I/O CONFIGURE screen’s
PCMCIA, Format Card field
The mass storage location for the built-in IBASIC controller can be
changed using the Mass Storage Is (MSI) command. Refer to the
Agilent Technologies Instrument BASIC Users Handbook for further
information on the MSI command.
Save/Recall Registers
Under the following conditions, the default mass storage location for
the save/recall registers is the Test Sets internal RAM (no mass storage
volume specifier):
power-up
initializing RAM with the SERVICE screen’s RAM Initialize
function
resetting the Test Set using the front-panel PRESET key
resetting the Test Set using the *RST HP-IB Common Command
The mass storage location for save/recall registers can be changed using
the Save/Recall field in the I/O CONFIGURE screen. The default
mass storage volume specifiers for the save/recall register mass storage
locations are as follows:
Internal selection - (no mass storage volume specifier, registers are
saved to allocated RAM space)
Card selection (not changeable) - ":INTERNAL,4"
RAM selection (not changeable) - ":MEMORY,0,0" see (“Using RAM
on page 221)