Memory Cards/Mass Storage
Default Mass Storage Locations
Default Mass Storage Locations
Built-in IBASIC Controller
Under the following conditions, the default mass storage location for
the built-in IBASIC controller is the front-panel card slot (mass storage
volume specifier ":INTERNAL,4"):
• power-up
• initializing RAM with the SERVICE screen’s RAM Initialize
function
• formatting an SRAM card using the I/O CONFIGURE screen’s
PCMCIA, Format Card field
The mass storage location for the built-in IBASIC controller can be
changed using the Mass Storage Is (MSI) command. Refer to the
Agilent Technologies Instrument BASIC Users Handbook for further
information on the MSI command.
Save/Recall Registers
Under the following conditions, the default mass storage location for
the save/recall registers is the Test Set’s internal RAM (no mass storage
volume specifier):
• power-up
• initializing RAM with the SERVICE screen’s RAM Initialize
function
• resetting the Test Set using the front-panel PRESET key
• resetting the Test Set using the *RST HP-IB Common Command
The mass storage location for save/recall registers can be changed using
the Save/Recall field in the I/O CONFIGURE screen. The default
mass storage volume specifiers for the save/recall register mass storage
locations are as follows:
• Internal selection - (no mass storage volume specifier, registers are
saved to allocated RAM space)
• Card selection (not changeable) - ":INTERNAL,4"
• RAM selection (not changeable) - ":MEMORY,0,0" see (“Using RAM”
on page 221)