User`s manual

Chapter 6 183
Advanced Operations
Instrument Initialization
The *RST Reset condition in the Test Set was specifically designed to
configure the instruments for manual testing of an FM radio. The *RST
Reset default display screen is the RX TEST screen. Other operational
characteristics are also affected by the *RST reset as follows:
All pending operations are aborted.
Measurement triggering is set to TRIG:MODE:SETT FULL;RETR
REP.
Any previously received Operation Complete command (*OPC) is
cleared.
Any previously received Operation Complete query command
(*OPC?) is cleared.
The Test Set’s display screen is in the UNLOCKED state.
The Power-up self-test diagnostics are not performed.
The Contents of the SAVE/RECALL registers are not affected.
Calibration data is not affected.
The HP-IB interface is not reset (any pending Service Request is not
cleared).
All Enable registers are unaffected: Service Request, Standard
Event, Communicate, Hardware #1, Hardware #2, Operation,
Calibration, and Questionable Data/Signal.
All Negative Transition Filter registers are unaffected:
Communicate, Hardware #1, Hardware #2, Operation, Calibration,
and Questionable Data/Signal.
All Positive Transition Filter registers are unaffected: Communicate,
Hardware #1, Hardware #2, Operation, Calibration, and
Questionable Data/Signal.
The contents of the RAM are unaffected.
The contents of the Output Queue are unaffected.
The contents of the Error Queue are unaffected.