User`s manual
Chapter 6 147
Advanced Operations
Status Reporting
Questionable Data/Signal Register Group
The Questionable Data/Signal Register Group contains information
about the quality of the Test Set’s output and measurement data. This
status group is accessed using the STATus commands. The
Questionable Data/Signal Register Group uses 16-bit registers and
includes a Condition Register, Transition Filters, an Event Register, an
Enable Register, and a Summary Message. Refer to the “Status
Reporting Structure Overview” on page 121 for a discussion of status
register operation. Figure 6-10 on page 147 shows the structure and
STATus commands for the Questionable Data/Signal Register Group.
Figure 6-10 Questionable Data/Signal Register Group
0
:STATus: QUEStionable: ENABle
?
<integer>
:STATus: QUEStionable: EVENt ?
:STATus: QUEStionable
:PTRansition
:NTRansition
?
<integer>
?
<integer>
:STATus: QUEStionable: CONDition ?
Summary
Message Bit
Condition
Register
Transition
Filter Registers
Event
Register
Enable
Register
1
2
3
4
5
6
7
8
9
10
11
12
13
14
L
o
g
i
c
a
l
O
R
15
ch4drw10.drw
(to bit 3 of Status
Byte Register)