User`s manual
Chapter 5 111
Triggering Measurements
Triggering Measurements
Trigger Mode Settings for Most Reliable Measurements
Use the following Test Set configuration and trigger mode settings to
get the most accurate, most reliable, fully settled measurement results.
See “Increasing Measurement Throughput” on page 114 for more
information on improving measurement throughput.
1. Turn on all autoranging functions. (This is the Test Set’s default
turn-on and preset state.)
2. Use SINGle RETRiggering.
3. Use FULL SETTling.
4. Individually trigger each measurement.
Measurement Pacing
Measurement pacing can be accomplished by using the IEEE
488.2-1987 Common Commands *OPC, *OPC?, and *WAI. These
commands are implemented within the Test Set using the criteria that
an operation has not completed until
• all active measurements have obtained at least one valid
measurement result
• all signals generated by the Test Set are within specifications.
Refer to the “Common Command Descriptions” on page 89 and the
IEEE 488.2-1987 Standard for more information on using these
commands.