User`s manual

108 Chapter 5
O:\Manuals\E6380A_Progguid\Book\Trigmeas.fm
Triggering Measurements
Triggering Measurements
Default Trigger Mode
The Trigger mode is set to FULL SETTling and REPetitive
RETRiggering whenever
the Test Set is powered on
the PRESET key is selected
the Test Set is put into LOCAL mode
the Test Set is reset using the *RST command
the Test Set is put in remote mode and no other trigger mode is set
Local/Remote Triggering Changes
Local To Remote Transitions
The Test Set switches from Local to Remote mode upon receipt of the
Remote message (REN bus line true and Test Set is addressed to
listen). No instrument settings are changed by the transition from
Local to Remote mode, but triggering is set to the state it was last set to
in Remote mode (if no previous setting, the default is FULL SETTling
and REPetitive RETRiggering).
When the Test Set makes a transition from local to remote mode, all
currently active measurements are flagged as invalid causing any
currently available measurement results to become unavailable. If the
HP-IB trigger mode is :RETR REP then a new measurement cycle is
started and measurement results will be available for all active
measurements when valid results have been obtained. If the HP-IB
trigger mode is :RETR SING then a measurement cycle must be started
by issuing a trigger event. Refer to Triggering Measurements” on page
105 for more information.
Remote To Local Transitions
The Test Set switches from Remote to Local mode upon receipt of the
Local message (Go To Local bus message is sent and Test Set is
addressed to listen) or receipt of the Clear Lockout/Set Local message
(REN bus line false). No instrument settings are changed by the
transition from Remote to Local mode, but triggering is reset to FULL
SETTling and REPetitive RETRiggering.