Specifications
Chapter 3 75
Troubleshooting
Manual Troubleshooting Procedures (Step 4)
Manual Troubleshooting Procedures (Step 4)
If you are not sure a problem exists, you should attempt to duplicate the
suspected problem. This is especially important if the Test Set is being
used in a new application where misapplication, or incorrect operation
of the Test Set may be involved.
An Agilent 8924C Mobile Station Test Set combined with an Agilent
83236B Cellular Adapter can be used to simulate a high performance
CDMA base station and may be useful in attempting to duplicate the
problem.
Refer to following table to determine which diagnostic tests,
performance tests, and periodic self calibration adjustments apply to an
assembly. Downloading calibration data is discussed in Chapter 7 ,
“Periodic Adjustments,” on page 157.
Table 3-2 Relating Assemblies to Troubleshooting Aids
Ref.
Designator
Assembly Name SERVICE4
Program
Diagnostic Test:
Sub-Test
Performance
Test to Perform
1
Periodic
Calibration
2
Program
Cal.-Data
Needed
3
A1A2 Keypad Functional
Diagnostics: Self
Test
No
A1A1 Display No
A2A200 100W Attenuator Functional
Diagnostics:
RF Modules
PCMCIA Program
System Power
E6380-61811
Yes
A2A130 RF Input/Output RF Diagnostics:
RF Input/Output
RF Generator:
Level Accuracy
PCMCIA Program
System Power
E6380-61811
Yes
A2A34 Data Buffer Functional
Diagnostics: CDMA
Loopback
SERVICE4: IQ
Modulator, Eb/No
Yes
A2A110 Upconverter RF Diagnostics:
Upconverter
Yes
A1A2 RPG Assembly No
A1 Front Panel No
A2A36 Receive DSP Functional
Diagnostics: CDMA
Loopback
No
A2A10 PCMCIA No