Troubleshooting
Functional Diagnostics (Step 2)
For descriptions of the diagnostic options, refer to:
“Functional Diagnostics Menu” on page 66.
AF Diagnostics” on page 68.
“RF Diagnostics” on page 70.
“CDMA Diagnostics” on page 72.
Define Test Conditions
1. On the SOFTWARE MENU screen (see Figure 3-11 on page 65),
select Execution Cond to access the TESTS (Execution Conditions)
screen.
2. Set up the Output Results To: field. Select:
Crt to view measurements only on the display.
Select Printer to print the test results as well as display them
on the CRT.
3. Set the Output Results For: field to All
4. Set up the If Unit-Under-Test Fails: field.
•Select Continue to continue to the next test point.
•Select Stop to pause testing at that point.
5. Set up the Test Procedure Run Mode: field.
•Select Continuous to run the tests continuously.
•Select Single Step to pause after each measurement.
6. Verify that the Autostart Test Procedure on Power-Up: setting
is Off.