Specifications
Chapter 3 63
Troubleshooting
Functional Diagnostics (Step 2)
For descriptions of the diagnostic options, refer to:
• “Functional Diagnostics Menu” on page 66.
• “AF Diagnostics” on page 68.
• “RF Diagnostics” on page 70.
• “CDMA Diagnostics” on page 72.
Define Test Conditions
1. On the SOFTWARE MENU screen (see Figure 3-11 on page 65),
select Execution Cond to access the TESTS (Execution Conditions)
screen.
2. Set up the Output Results To: field. Select:
• Crt to view measurements only on the display.
• Select Printer to print the test results as well as display them
on the CRT.
3. Set the Output Results For: field to All
4. Set up the If Unit-Under-Test Fails: field.
•Select Continue to continue to the next test point.
•Select Stop to pause testing at that point.
5. Set up the Test Procedure Run Mode: field.
•Select Continuous to run the tests continuously.
•Select Single Step to pause after each measurement.
6. Verify that the Autostart Test Procedure on Power-Up: setting
is Off.