Specifications
264 Chapter 9
Performance Test Records
AF Analyzer Distortion and SINAD Accuracy Performance Test 14 Record
AF Analyzer Distortion and SINAD Accuracy
Performance Test 14 Record
For test procedure, see “AF Analyzer Distortion and SINAD Accuracy
Performance Test 14” on page 203.
Table 9-14 AF Analyzer Distortion and SINAD Accuracy Test 14 Record
AF
Generator 2
Frequency
(kHz)
AF Generator
2 Level (mV)
Measurement
Type
Distortion and SINAD Limits
Lower Upper Actual
2 100 Distortion 8.856% 11.144%
2 100 SINAD 19.043 dB 21.043 dB
3 100 Distortion 8.856% 11.144%
3 100 SINAD 19.043 dB 21.043 dB
2 10 Distortion 0.890% 1.120%
2 10 SINAD 39.000 dB 41.000 dB
3 10 Distortion 0.890% 1.120%
3 10 SINAD 39.000 dB 41.000 dB
2 5 Distortion 0.445% 0.560%
2 5 SINAD 45.021 dB 47.021 dB
3 5 Distortion 0.445% 0.560%
3 5 SINAD 45.021 dB 47.021 dB