Specifications
Chapter 8 227
Performance Tests
CDMA Analyzer Channel Power Level Accuracy Performance Test 28
CDMA Analyzer Channel Power Level Accuracy
Performance Test 28
The purpose of this test is to verify that the Test Set meets the
specification limits in PTR (Performance Test Record) Table 9-28,
“CDMA Analyzer Average Power Level Accuracy Test 27 Record,” on
page 279. The tuned channel power level accuracy is verified by
comparing the measured power in a CW signal with the power level
measured by a measuring receiver.
Setup
Figure 8-33 CDMA Analyzer Channel Power Level Accuracy Test 28
Procedure
1. On the measuring receiver
a. Set the display mode to LOG.
b. Set the measurement mode to RF Level.
c. Calibrate the power sensor.
Measuring Receiver
INPUT 50Ω
SENSOR
RF IN/OUT
Amplifier 2
Signal Generator
OUTPUT
Sensor Module
Power Splitter