Specifications
Chapter 8 225
Performance Tests
CDMA Analyzer Average Power Level Accuracy Performance Test 27
CDMA Analyzer Average Power Level Accuracy
Performance Test 27
The purpose of this test is to verify that the Test Set meets the
specification limits in PTR (Performance Test Record) Table 9-27,
“CDMA Generator Modulation Accuracy Test 26 Record,” on page 278.
The CDMA average-power-level accuracy is verified by comparing the
measured power in a CW signal with the power level measured by a
power meter.
Setup
Figure 8-32 CDMA Analyzer Average Power Level Accuracy Test 27
Procedure
1. On the measuring receiver:
a. Set the measurement mode to RF Power.
b. Calibrate the power sensor.
2. On the signal generator:
a. Set the frequency to 881.32 MHz.
b. Set the amplitude so the measuring receiver reads 4 mW.
Measuring Receiver
INPUT 50Ω
SENSOR
RF IN/OUT
Amplifier 2
Signal Generator
OUTPUT
Sensor Module
Power Splitter