signal comes from the external audio source in the audio analyzer. The
audio level is varied until the modulation is at the desired FM deviation
as measured by the measuring receiver.
NOTE
Use the AM/FM test source output labeled FM÷32 for 12.5 MHz and the
output labeled FM for 400 MHz. You can measure the frequency with
the measuring receiver and adjust it with the carrier frequency tune
knob, but the exact frequency is not critical.
Figure 8-25 Setup
Procedure
1. On the AM/FM test source, set the test mode to FM.
2. On the measuring receiver:
a. Reset the instrument.
b. Set the measurement mode to FM.
c. Set the detector to RMS.
ANT IN
Audio Analyzer
Measuring Receiver
AM/FM Test Source
FM÷32 or FM OUTPUT
AUDIO
INPUT
OUTPUT
INPUT 50