Specifications
186 Chapter 8
Performance Tests
RF Generator FM Flatness Performance Test 3
5. On the measuring receiver:
a. Set the frequency offset mode to enter and enable the LO
frequency (27.3 Special).
b. Key in the LO frequency (in MHz) which is 1500.
6. On the Test Set, for frequencies of 1700 and 2000 MHz, measure the
FM deviation at the rates shown in the PTR. Convert the
measurement results as was done in step 3 and compare the
calculated deviation to the limits.