Specifications
Chapter 7 163
Periodic Adjustments
Running the Periodic, IQ, or Eb/No Calibration Programs
Running the Periodic, IQ, or Eb/No
Calibration Programs
1. Press Menu to access the SOFTWARE MENU screen.
2. Select the field under Select Procedure Location:.
3. Select ROM under the Choices: menu.
4. Select the field under Select Procedure Filename:.
5. Select SERVICE4, see Figure 7-2 on page 163.
6. Select Run Test (key
k1).
7. From the SERVICE MENU, select the desired calibration program
to perform.
• Periodic Calibration - for more detailed information, see
“Periodic Calibration Menu Descriptions” on page 165.
• IQ Calibration - for more detailed information, see “IQ
Calibration Program Description” on page 170.
• Eb/No Calibration - for more detailed information, see “Eb/No
Calibration Program Description” on page 171.
8. Follow the instructions on the screen.
Figure 7-2 SERVICE MENU Screen
servsrn1.eps
SOFTWARE MENU Screen
SOFTWARE MENU
Description:
Launches disgnostic and calibration programs.
Continue
1
2
4
Run Test
Help
SOFTWARE MENU
Library:
[NO LIB]
Program:
ROM
CUSTOMIZE TEST PROCEDURE:
Channel Information
Test Parameters
Order of Tests
Pass/Fail Limits
Save/Delete Procedure
SET UP TEST SET:
Execution Cond
External Devices
Printer Setup
IBASIC Cntrl
LOAD TEST PROCEDURE:
Select Procedure Location:
ROM
Select Procedure Filename:
SERVICE4
Exec
Cnfg
Print
IBASIC
Freq
Parm
Seqn
Spec
Proc
SERVICE MENU
1
2
3
4
5
Help
Exit
Functional Diagnostics
AF Diagnostics
RF Diagnostics
CDMA Diagnostics
Edit RF Diagnostic Limits
Periodic Calibration
IQ Calibration
Eb/No Calibration
=>
Move pointer to the desired program using the
knob then press the knob. Press Help for
information on the tests. Press Exit to abort.
servscn2.eps
SERVICE MENU Screen