Specifications
136 Chapter 6
Replaceable Parts
Replacement & Ordering Parts
Replacement & Ordering Parts
Direct Parts Ordering
See “Factory Support” on page 46. The Agilent Support Materials
Organization can help you order and identify parts.
Assembly Replacements
With some assemblies you will receive a Memory Card that contains
factory-generated calibration data for that assembly. There will also be
an instruction sheet for loading the calibration data into your Test Set
after you’ve replaced the assembly.
NOTE
Periodic Adjustment Interval
The adjustment programs Periodic Calibration, IQ Calibration, and
Eb/No Calibration should be performed after any assembly referred to
in Table 7-1 on page 159 is replaced, or at least every 24 months. These
program can be run anytime to optimize the performance of the Test
Set. See Chapter 7 , “Periodic Adjustments,” on page 157 for details.
NOTE
Performance Test Interval
The performance tests in Chapter 8 , “Performance Tests,” on page 173
should be performed when certain assemblies are repaired or replaced,
or at least every 24 months. See Table 3-2 on page 75 for those
assemblies requiring performance testing/calibration.