User`s guide

51
System Specifications
1710 to 1990 MHz Aux RF Out Connector Specifications (HP/Agilent 83220A only)
1710 to 1990 MHz Aux RF Out Connector Specifications
(HP/Agilent 83220A only)
1
Frequency
Range: 1710 MHz to 1990 MHz
Resolution: 1 Hz
Accuracy: Reference accuracy ± 0.5 Hz
Stability: Same as reference
Output
Level Range for specified accuracy: +7 to -127 dBm
Level resolution: 0.1 dB
Level accuracy: ± 1 dB (1710 to 1880 MHz)
± 1 dB typical (1880 to 1990 MHz)
Reverse power: 200 mW
SWR: 2.0:1 for level <-4 dBm
Spectral purity
Spurious signals: (for +1 dBm output level at Aux RF
out, or -19 dBm output level at RF in/
out)
Harmonics: < -25 dBc
Non-harmonics: < -50 dBc, > 5 kHz offset from carrier
1. HP/Agilent 8922X and HP/Agilent 8922Y systems only