Screens
Dual Band Control Screen
6. Execute
This field starts the process of moving the traffic channel from one band to
another. If no call is active, the inter-band channel change cannot be executed. If
the inter-band channel change fails or the call drops for any other reason, the
operating mode defaults to E-GSM900.
7. Input Level
This field sets the expected input level of the HP/Agilent 8922 RF Analyzer. The
value set in this field will only be used once a inter-band channel change has been
executed.
This field will be set by the
TX Level field, if the Level Control field is set to
.
The Level Control field will change to if the Input Level is
changed.
When in compatibility mode (8922E or 8922G) and Level Control field is set
to , then an error will occur if the
Input Level is changed.
8. Level Control
This field couples the TX Level and Input Level fields.
This field will be set to if the
Input Level is altered.
MS TX Lev
Manual
MS TX Lev
E-GSM900 Default 13 dBm
Range -27.9 dBm to 41.0 dBm
DCS1800 Default 10 dBm
Range -47.9 dBm to 33.0 dBm
Manual
Choices Any changes to the TX Level field will also
change the value of the Input Level field.
Any changes to the TX Level field will effect
the Input Level field and the
Level Control changes automatically to
MS TX Lev
Manual
MS TX Lev