User`s guide
10
Configuring the HP/Agilent 8922 Multi-Band Test System
Zero the Power Meter
4 Once selected, the power meter will automatically zero.
To Zero the Power Meter of the HP/Agilent 83220
NOTE: In Dual-Band test mode the Test Set is initially set to E-GSM and in this mode the power
zero is applied only to the HP/Agilent 8922 Power Meter. Because of this, the HP/Agilent
83220 Power Meter must be zeroed separately by setting the Radio Type to DCS1800.
1 Ensure that no external instruments, test devices or mobiles are connected to the HP/
Agilent 8922 Multi-Band Test System.
2 Switch on the HP/Agilent 8922 Multi-Band Test System, navigate to the CONFIGURE
screen, select the Radio Type and set it to DCS1800 or PCS1900.
3 Press the front panel to display the Cell Status screen and select the
field.
4 Once selected, the power meter will automatically zero.
For details of recommended GPIB commands for including power zeroing in the
initialization routine for testing of Dual-Band type mobiles, see “GPIB
Commands for Including Power Zeroing in an Automatic Test Routine” on
page 4-44.
CELL
CNTL
Pwr Zero
Power Zero