User`s guide
7
Configuring the HP/Agilent 8922 Multi-Band Test System
Calibration Routine
Calibration Routine
Once the HP/Agilent 8922 Multi-Band Test System has been connected, the
system must be calibrated. The calibration routine can be used for both the HP/
Agilent 83220A and the HP/Agilent 83220E.
NOTE: The calibration routine must be run whenever one or more of the following situations arise:
❒ An HP/Agilent 83220 is first connected to an HP/Agilent 8922.
❒ When a different, or upgraded, HP/Agilent 83220 is connected to an HP/Agilent 8922.
❒ If different IF Link or Aux RF In/Out cables are used.
❒ After a RAM Initialize or Firmware Upgrade.
❒ Need to trouble shoot the system or localize faults.
Procedure
1 Ensure you have the following parts
1
:
• 6 dB coaxial attenuator (SMA)
• BNC to SMA adapter
• N-type to SMA adapter
• BNC to BNC cable (minimum length 1.2m)
2 Ensure that the Coupled RF In/Out Port is terminated by 50 Ω (supplied with test
system).
3 Switch on the HP/Agilent 8922 Multi-Band Test System.
4 Select from the To Screen.
5 Select Compatible field and change to or , if required.
6 Power cycle the system.
7 Press the front panel key.
8 Set Location to ROM
9 Select Procedure field.
10 Select the test routine SYS_CAL.
11 Run the program by either selecting the Run Test field or by pressing .
1. These are a list of parts included in the calibration kit - HP/Agilent 8922U Option 104.
CONFIG
8922E 8922G
TESTS
L1