Specifications
Chapter 4 76
Tests
TEST_02 CP BS Originate
TEST_02 CP BS Originate
Description This test verifies that a mobile station can be paged by simulating a
base station originated call.
Entering the IMSI number into the parameters list is optional. If the
default IMSI (all 0’s) is detected, the software prompts you to make a
mobile station originated call. The IMSI is then obtained by the
HP/Agilent 8922 and a base station originated call is attempted. If an
incorrect IMSI is entered, a base station originated call is unsuccessful.
The following information is displayed when the test is run:
• Pages until mobile is connected
• RACHs until mobile is connected
• Call-processing timing advance
• Call-processing mobile station TX level
• Call-processing trigger-timing error
Specifications None
Parameters
Parm
no.
Description Units GSM900
default
value
DCS1800
default
value
PCS1900
default
value
1 AE IMSI (digits 1-5) number 0.0 0.0 0.0
2 AE IMSI number (digits 8-15) number 0.0 0.0 0.0
3 CP base station color code code no. 5.0 5.0 5.0
4 CP public land mobile network color code code no. 1.0 1.0 1.0
5 CP serving cell (BCH) ARFCN chan. no. 25.0 600.0 600.0
6 CP local area code code no. 1.0 1.0 1.0
7 CP mobile country code code no. 1.0 1.0 1.0
8 CP mobile network code code no. 1.0 1.0 1.0
9 CP control chan type [0=SD/8, 1=SD/4] 0 or 1 1.0 1.0 1.0
10 CP test with ciphering [0=no 1=yes] 0 or 1 0.0 0.0 0.0
11 CP delay(+) or advance (-) for trigger T 0.0 0.0 0.0
12 RT default traffic channel [ARFCN 1-124] chan. no. 62.0 699.0 699.0
14 RT TCH [0=static 1=hopping 3 freqs] 0 or 1 0.0 0.0 0.0
16 RT external pad and cable loss dB 0.0 0.0 0.0