Specifications
Chapter 4 74
Tests
TEST_01 MS Information
TEST_01 MS Information
Description This test prompts you to make a mobile station originated call, and the
IMSI number is acquired by the HP/Agilent 8922. At the beginning of
the test you also have the option to enter the mobile’s serial number.
The following MS Information is displayed:
• GSM Phase
• Which bands the mobile supports
• MS Power Class
• MS IMSI (International Mobile Subscriber Identity Number)
• MS IMEI (International Mobile Equipment Identification)
• MS ONUM (Originated Number)
• Whether or not the MS has passed an IMEI consistency check. The
read IMEI must be in the format of 15 decimal digits (no Hex) and
the final character must be 0. The message IMEI check
successfully passed or IMEI does not fulfill
requirements appears depending on the outcome of the check.
Specifications None
Parameters The following parameters are used in TEST_01.
Parm
no.
Description Units GSM900
default
value
DCS1800
default
value
PCS1900
default
value
1 AE IMSI (digits 1-5) number 0.0 0.0 0.0
2 AE IMSI number (digits 8-15) number 0.0 0.0 0.0
3 CP base station color code code no. 5.0 5.0 5.0
4 CP public land mobile network color code code no. 1.0 1.0 1.0
5 CP serving cell (BCH) ARFCN chan. no. 25.0 600.0 600.0
6 CP local area code code no. 1.0 1.0 1.0
7 CP mobile country code code no. 1.0 1.0 1.0
8 CP mobile network code code no. 1.0 1.0 1.0
9 CP control chan type [0=SD/8, 1=SD/4] 0 or 1 1.0 1.0 1.0
10 CP test with ciphering [0=no 1=yes] 0 or 1 0.0 0.0 0.0
11 CP delay(+) or advance(-) for trigger T 0.0 0.0 0.0
12 RT default traffic channel [ARFCN 1-124] chan. no. 62.0 699.0 699.0
14 RT TCH [0=static 1=hopping 3 freqs] 0 or 1 0.0 0.0 0.0
16 RT external pad and cable loss dB 0.0 0.0 0.0