Specifications

Chapter 2 53
Using Worksheets
WORKSHEETS FOR DCS1800
NOTE
Each time a procedure is selected, the specifications list assumes the
values defined by that procedure.
22 TX power/time template ±18µs dBc or
dBm
30 dBc or
20 dBm
n/a n/a
23 TX power/time template
±28µs
1
dBc or
dBm
48 dBc or
48 dBm
n/a n/a
24 TX power/time template neg
peak flatness
dB 1.00
25 TX power/time template pos
peak flatness
dB 1.00
26 TX SACCH RX level error dB 3.00 3.00
27 TX RACH burst timing
measurement
T 1.00 1.00
28 TX RACH frequency error Hz 180.00 180.00
29 TX RACH peak power error dB
44
n/a n/a
30 TX RACH phase error (peak) degrees 20.00
31 TX RACH phase error (rms) degrees 5.00
32 TX RACH power/time
template at ±10 µs
dB
6.00
3
n/a n/a
33 TX RACH power/time
template at ±18 µs
dB 30.00 n/a n/a
34 TX RACH power/time neg
peak flatness
dB 1.00
35 TXRACHpower/time pos peak
flatness
dB 1.00
36 TX worst case frequency error Hz 180.00 180.00
37 TX worst case phase error
(peak)
degrees 20.00
38 TX worst case phase error
(rms)
degrees 5.00
39 TX peak power error for pcl 9dB
44
n/a n/a
1 HP/Agilent 8922 with Spectrum Analyzer Option 006 only
2 PCL (power Control Level) refers to mobile TX level
3 For power control levels 16, 17 and 18 and above see page 123
4 Dependent on power control level, see page 123
DCS1800 SPECIFICATIONS worksheet for:
Used? Step
no.
Description Units DCS1800
default
lower
limit
DCS1800
default
upper
limit
Lower
limit
Upper
limit