Making your own test procedures
Making your own test procedures
Quick overview To make a test procedure may require some understanding of HP/
Agilent 8922 operation. If necessary, refer to the HP/Agilent 8922
User’s Guide. An overview of the process for making your own test
procedure is shown below.
NOTE
To create your own Dual-Band procedure, or change any of the
parameter values, you must use a separate memory card. The
procedure must be called DUALBAND and the procedures for GSM900
and DCS1800 must also exist on the memory card to provide the
Dual-Band procedure with the necessary parameters and specifications
for each band.
Figure 1-10 Overview of making a test procedure
Before making a Test Procedure
* Run a pre-defined test procedure
* Edit the test procedure
* Select Edit Proc from Seq Menu
Edit the Test Sequence
* Select Edit Seqn from TESTS screen
* Modify existing TEST sequence
* Insert/delete tests
Edit the Procedure PARM
* Select Edit Parm from TESTS screen
* Modify existing parameters
* Insert/delete tests
Edit the Procedure SPEC
* Select Edit Spec from TESTS screen
* Modify existing specifications
* Insert/delete tests
Verify the modified Procedure will run
* Connect a mobile
* From the TESTS screen, press L2 to return to the Edit Proc screen
* Press L2 to return to the modified procedure
Save the Test Procedure to an SRAM card
* Press Cancel followed by TESTS
* Select Proc Mngr screen from the Test Function field
* Enter a Procedure name and comment
* Insert and initialized SRAM card
* Select Make Procedure to save the procedure to the card