Specifications
Chapter 1 23
Operating the Agilent Technologies 83212D Software
Selecting a test sequence — Test Sequence menu
Test execution conditions
On the lower half of the Sequence Menu there are a number of toggle
fields which affect test operation:
• Auto-repeat ON/OFF : When this field is set to ON, the selected
test sequence will run repeatedly WITHOUT prompting you. However,
an ABORT softkey will be available at various times
throughout testing.
• Stop on failure YES/NO : When this field is set to YES, testing
will stop whenever the mobile fails to meet a test specification limit
(defined by the selected procedure). You will then be prompted to
continue the test.
• Results output ALL/FAILS : When this field is set to ALL, all test
results are sent to the HP/Agilent 8922 display and printer (if
enabled). When this field is set to FAILS, only the following are sent
to the HP/Agilent 8922 display and printer (if enabled):
— Test results which fail to meet specification limits (which are
defined by the selected procedure).
— Testing errors (for example, failure of a mobile to respond).
• Printer ON/OFF : When this field is set to ON, test results which
are displayed on the HP/Agilent 8922 screen will also be output to a
printer. The printer first must be set up using the Printer
Configuration Menu (see “Printing test results” on page 35 for
further details.)
• Run mode CONT/STEP : When this field is set to STEP, the program
will step through the test, pausing after each result. Each time the
user will be prompted to continue testing by pressing .
When this field is set to CONT, the program does not pause after each
result.
• Print Procedure : Selecting this option allows a listing of the test
sequence, specifications and parameters of the currently selected
procedure to be sent to a printer. The printer first must be set up
using the Printer Configuration Menu as described in “Printing
test results” on page 35.
L2
L2