Specifications

Chapter 1 9
Operating the Agilent Technologies 83212D Software
Getting started
Upgrading to a new software revision
NOTE
It is essential when loading the new revision of the Agilent 83212D for
the first time that you remove any previous revision, otherwise the new
software does not load correctly. The simplest way to remove the old
revision is to run an IBASIC application from ROM as follows.
Ignore the instructions that appear on the HP/Agilent 8922 display
during the procedure.
Step 1. Go to the screen and select the field named Location.
Step 2. From the list of choices, select ROM.
Step 3. Select the preceding field Procedure.
Step 4. Select RAM_MNG (RAM Manager).
Step 5. Select or select Run Test which will cause the selected program
to overwrite any existing code.
Step 6. Select or select No in reply to the question Do you want to
continue?
Step 7. Return to the screen and load up the Agilent 83212D
software from the memory card as described in “Loading the Agilent
83212D software from memory card” on page 8.