Specifications
Chapter 7 177
Screens
TESTS (Procedure Manager)
5. Edit Seqn This field is used to select tests and test sequence.
6. Edit Spec This field is used to enter test limits.
7. Library For
New Procedure
Choices are:
• Current selects use of the current library file with the test
procedure.
• NO LIB will cause test procedures to be created with no associated
library file.
8. Location to
Make or Delete
Choices are:
• Card selects the MEMORY CARD slot on the front panel of the
HP/Agilent 8922.
• RAM selects internal RAM.
• Disk selects and external disk drive.
9. Pass Number This field is not used for by the Agilent 83212D software.
10. Procedure This field is used to compose the name of a test procedure.
Step 1. Select the Procedure field. (An alpha/numeric list of characters
appears in the lower-right corner of the screen.)
Step 2. Use the knob on the front panel of the HP/Agilent 8922 to select
characters in order to compose the procedure name.
Step 3. Select Done when you are finished.
11. Program
Location For
New Procedure
This field selects the location for the Agilent 83212D program file when
a test is run. For Agilent 83212D operation, it is recommended that this
selection be CARD. Choices are:
• Card: selects the memory card in the MEMORY CARD slot on the
front panel of the HP/Agilent 8922 as the location for the Agilent
83212D program file.
• ROM: selects internal ROM.
• RAM: selects internal RAM.
• Disk: selects an external disk drive.
12. Test Function This field selects the screens used in the tests subsystem. Choices are:
• Edit Seqn is used to select tests and test sequence.
• Edit Freq is not used for by the Agilent 83212D software.
• Edit Spec is used to enter test limits.
• Edit Parm is used to enter test parameters.