Specifications
Chapter 7 166
Screens
TESTS
11. Output
Results
Choices are:
• All: when All is selected, all test results are shown on the CRT
and/or printout.
To send data to a printer see “Printing test results” on page 35.
• Failures: when Failures is selected, test results are shown only
when a UUT fails to meet test specification limits or when a software
error occurs.
12. Procedure This field selects and displays a test procedure or program from the
Choices area in the lower right portion of the display. To display a list
of choices, choose CARD or ROM in the Location field, then select the
Procedure field.
NOTE
Selecting procedures in this field loads a new procedure, causing
existing sequence, specifications, and parameters to be overwritten.
13. Program This field displays the location of the program file.
14. Run Mode Choices are:
• Continuous: in Continuous Mode, all tests run in sequence. The
test will pause only when the operator is required to interact with
the UUT or HP/Agilent 8922, such as when entering a phone
number on the mobile station, or when the CANCEL key is pressed or
the Abort softkey is selected.
• Single Step: in Single Step Mode, the program pauses at the
completion of each measurement in the test procedure. The
test-system operator is prompted to select Continue to proceed with
testing.
15. Run Test This field begins running Agilent 83212D software. If the Agilent
83212D software has not been loaded, selecting this field will display
“Loading program - Enter SHIFT-CANCEL to abort”. The software
takes about two minutes to load. The loading operation is performed
the first time the program is run. The software does not have to be
re-loaded unless a ROM program is run or RAM is cleared.