Specifications
Chapter 7 165
Screens
TESTS
7. Location This field selects the location of programs, test procedures, and files.
Choices are:
• Card selects a memory card inserted in the MEMORY CARD slot in
the HP/Agilent 8922 front panel.
• ROM selects internal ROM. Several programs will be displayed under
Choices in the lower right portion of the display. A program is
selected using the Procedure field; select Run Test to run the
program.
• RAM selects internal RAM.
• Disk selects an external disk drive.
8. On UUT
Failure
Choices are:
• Continue: when Continue is selected, testing continues even when
the mobile fails to meet its test specification limits.
• Stop: When Stop is selected, testing stops whenever the mobile
fails to meet test specification limits.
Note that test fields 8, 9, 11, and 14 are replicated on the Sequence
menu. Hence, it is not necessary to set these fields on this screen.
9. Output
Destination
Choices are:
• Crt: when Crt is selected, test results are output to the HP/Agilent
8922 CRT screen only.
• Printer: When Printer is selected, test results are output to the
CRT and the system printer. The printer must be correctly
configured to get a printout—see “Printing test results” on page 35.
10. Output
Heading
This field is used to enter a heading which will appear with test results,
both on the CRT display and a printout.
To enter an output heading:
Step 1. Select the Output Heading field (an alpha/numeric list of characters
appears in the lower-right corner of the screen.)
Step 2. Use the knob on the front panel of the HP/Agilent 8922 to select
characters (two lines of comments, 50 characters in length, may be
entered.)
Step 3. Select Done when you are finished.